The HZ-3015A hydrolysis resistance test chamber simulates controlled temperature and humidity environments for accelerated conditioning of footwear materials, rubber, plastics, electronics, automotive parts and metals.
Applicable standards: GB/T 2423 and GJB 150A1/4.
Technical specifications:
1. Resolution: temperature ±0.01°C; humidity ±0.1% RH.
2. Temperature range: ambient to +150°C.
3. Temperature fluctuation: ±0.2°C; uniformity: ±0.5°C.
4. Heating rate: 4–7°C/min, nonlinear and unloaded.
5. Cooling rate: 0.7–1.5°C/min, nonlinear and unloaded.
6. Humidity range: 20%–98% RH.
7. Humidity fluctuation/deviation: ±2% RH.
8. Internal dimensions: 400 × 400 × 500 mm.
9. External dimensions: 870 × 1,050 × 1,700 mm.